LIGO Production Large Actuator Test Report |
Actuator Serial #: L028
Operator Name: Smith, Lane
Date: 8/11/2009 Time: 2:03 PM
Actuator Coil Resistance: 6.39 Ohms, PASS
Ambient Temperature: 75.4 F
Hi Pot Test Results: 1000 MOhms, PASS
X Travel Limit (inches): 0.521
Y Travel Limit (inches): 0.191
Z Travel Limit (inches): 0.480
|
X Position (in) |
Y Position (in) |
Z Position (in) |
Current (A) |
Force (lbs) |
Force Constant (lbs/A) |
PASS/FAIL |
Center |
0.000 |
0.000 |
0.000 |
2.181 |
20.461 |
9.383 |
PASS |
Position 1 |
0.250 |
0.075 |
0.215 |
2.460 |
20.242 |
8.227 |
PASS |
Position 2 |
0.250 |
-0.075 |
0.215 |
2.312 |
19.855 |
8.586 |
PASS |
Position 3 |
-0.250 |
-0.075 |
0.215 |
2.367 |
19.914 |
8.411 |
PASS |
Position 4 |
-0.250 |
0.075 |
0.215 |
2.493 |
20.407 |
8.186 |
PASS |
Position 5 |
-0.250 |
0.075 |
-0.215 |
2.396 |
20.078 |
8.379 |
PASS |
Position 6 |
-0.250 |
-0.075 |
-0.215 |
2.364 |
19.966 |
8.446 |
PASS |
Position 7 |
0.250 |
-0.075 |
-0.215 |
2.360 |
20.015 |
8.480 |
PASS |
Position 8 |
0.250 |
0.075 |
-0.215 |
2.382 |
20.180 |
8.473 |
PASS |
Center |
0.000 |
0.000 |
0.000 |
-2.113 |
-20.060 |
9.494 |
PASS |
Position 1 |
0.250 |
0.075 |
0.215 |
-2.416 |
-20.578 |
8.518 |
PASS |
Position 2 |
0.250 |
-0.075 |
0.215 |
-2.308 |
-19.937 |
8.639 |
PASS |
Position 3 |
-0.250 |
-0.075 |
0.215 |
-2.302 |
-20.014 |
8.692 |
PASS |
Position 4 |
-0.250 |
0.075 |
0.215 |
-2.347 |
-20.634 |
8.792 |
PASS |
Position 5 |
-0.250 |
0.075 |
-0.215 |
-2.674 |
-20.015 |
7.486 |
PASS |
Position 6 |
-0.250 |
-0.075 |
-0.215 |
-2.794 |
-19.919 |
7.128 |
PASS |
Position 7 |
0.250 |
-0.075 |
-0.215 |
-2.498 |
-20.323 |
8.135 |
PASS |
Position 8 |
0.250 |
0.075 |
-0.215 |
-2.583 |
-19.985 |
7.736 |
PASS |
QNA-TSG
1901 South Harbor City Blvd
Suite 700
Melbourne, FL 32901
Contact: Scott Stewart
scott.stewart@plansys.com
321-768-6500 x11
© 2009 QinetiQ North America, Technology Solutions Group |