2013-07-10 08:44:57 jay.copti@LIGO.ORG 05/31/2013 - CIT - Build (SRI Hermetics Inc.) 05/31/2013 Jay Copti Modification Description The following components were modified for Notch Adjustment Frequencies RFPD (36 & 45 ) Cap/Res Inductors C6: 10 PF L1: 750 nH C9: 10 PF L2: 100 nH C10: 20 PF L3: 1800 nH C14: 10 PF L4: Remove C16: 20 PF L5: 100 nH C21: 40 PF (Var) L6:100 nH C22: 30 PF (Var) L7: 470 nH C23: 20 PF (Var) L8: 100 nH C24: 20 PF (Var) R9: 200 Ohms 05/31/2013 Jay Copti Acceptance Test Description and Results Acceptance test per Test Plan DCC # T1200347-v1 and a list of ISC detector measured parameters (RF & DC) data recorded can be found in T1200346-v3 under S/N: S1300632 05/31/2013 Jay Copti Manufacturing Notes It was shipped to Fred Williams at SRI Hermetics Inc. to install the back cover and seal it 2013-07-10 08:53:35 jay.copti@LIGO.ORG 07/09/2013 - CIT - Test (N.A.) 07/10/2013 Jay Copti Acceptance Test Description and Results Acceptance test per test plane Dcc # T1300506-v1 in clean room environment test was good it pass 2013-07-10 09:43:39 charles.osthelder@LIGO.ORG 07/10/2013 - LLO - Transfer (shipped to Valera Frolov, LLO) 07/10/2013 Charles Osthelder Manufacturing Notes Class A clean, ventilated QPD installed. Shipped with Class A clean base, alumina spacer, PEEK washers and vented SS socket cap screws, 1/4"-20x 3/4" and 4-40x3/8" for final assembly at site.